NABL ACCREDITED · ISO/IEC 17025:2017

6 Ways We Prove Your Device Delivers Every Breath Right

Respiratory and Aerosol Device Testing for nebulizers, ventilator circuits, and inhalation accessories at our Oragadam laboratory — airflow resistance, leakage, pressure drop, nebulizer output, and particle size evaluation referenced to EN 13544-1 and ISO 23328-1.

Call 90876 86986 WhatsApp Us

A nebulizer or ventilator circuit that looks fine on the shelf can still starve a patient of medication, leak breathable pressure into the room, or deliver aerosol particles too large to ever reach the lower airway.

Respiratory and Aerosol Device Testing answers the question a visual inspection never can: does the device move air with the resistance it claims, does it hold pressure without leaking, and does the medication it aerosolizes actually reach the part of the lung it’s meant to treat. Our Respiratory and Aerosol Device Testing programme at the Oragadam laboratory covers six checkpoints across connector integrity, airflow, pressure, and aerosol performance, referenced to EN 13544-1 for nebulizing systems, ISO 23328-1 for breathing system filters, ISO 80601-2-12 for ventilators, and USP <601> for aerosol particle size.

Connector integrity testing confirms every fitting in the breathing circuit seals correctly and resists dye penetration under pressure, because a single loose connection anywhere along the circuit undermines every downstream measurement of flow and pressure. Airflow resistance testing for medical device breathing circuits measures how much resistance a patient encounters when inhaling and exhaling through the device on a calibrated breathing simulator head, confirming the resistance stays within the range a compromised patient can actually tolerate. Leakage testing pressurizes the assembled circuit to confirm no breathable gas escapes anywhere along its length, since a leak in a ventilator circuit means the delivered volume never matches the volume the machine reports.

Pressure drop testing quantifies the pressure lost across filters, valves, and tubing under simulated breathing flow, because excessive pressure drop forces a patient to work harder for every breath even when the device otherwise passes every other check. Nebulizer output testing for medical device inhalation systems measures how much medication actually leaves the device per minute, confirming the dose a clinician calculates on paper is the dose a patient actually receives, and aerosol particle size distribution testing for medical device nebulizers characterizes the droplet size spread produced, since particles outside the respirable range simply deposit in the mouth and throat instead of reaching the lungs. Manufacturers bring their programmes to Kiyo R&D Lab because a single missed checkpoint in Medical Device Testing doesn’t surface as a bench-test footnote — it surfaces as an under-dosed patient, a leaking ventilator circuit, or medication that never reaches the airway it was meant to treat. Running all six checkpoints under one accredited scope at Oragadam, with reports formatted for direct regulatory submission, is what turns respiratory device testing into evidence your quality team can actually stand behind.

RESPIRATORY & INHALATION EVALUATION

Six Accredited Respiratory Device Checkpoints

Part of our Respiratory and Aerosol Device Testing panel, referenced to EN 13544-1 and ISO 23328-1

Connector integrity dye penetration test for respiratory medical device
01

Connector Integrity Test

ISO 5356-1

Airflow resistance testing with breathing simulator head for medical device
02

Airflow Resistance Test

ISO 23328-1

Respiratory and aerosol device leakage testing on ventilator for medical device
03

Leakage Test

ISO 80601-2-12

Pressure drop testing instrument for respiratory medical device filter
04

Pressure Drop Test

ISO 23328-1

Nebulizer output testing kit with mask for medical device
05

Nebulizer Output Test

EN 13544-1

Aerosol particle size distribution testing instrument for medical device
06

Aerosol Particle Size Distribution Test

USP <601>

How Our Respiratory Device Testing Process Works

Four steps, from sample intake to a validated report

1

Sample Intake

Device received against your target EN 13544-1 or ISO 23328-1 sub-clause.

2

Connector & Leakage

Circuit fittings and full assembly checked for seal integrity under pressure.

3

Airflow & Aerosol

Airflow resistance, pressure drop, nebulizer output, and particle size measured.

4

Certified Report

Delivered digitally, ready for your validation file.

Standards Behind Our Medical Device Testing

CHECKPOINT SCOPE STANDARD
Airflow resistance testing for medical deviceBreathing resistance on calibrated simulator headISO 23328-1
Nebulizer output testing for medical deviceMedication mass delivered per minuteEN 13544-1
Aerosol particle size distribution testing for medical deviceDroplet size spread for respirable fractionUSP <601>
Leakage TestCircuit seal integrity under pressureISO 80601-2-12
VISIT OUR LABORATORY

Kiyo R&D Lab – Oragadam

Our respiratory and aerosol device testing suite is based at No.13, Srinivasa Perumal Street, Panapakkam, Serapanacheri, Greater Chennai, Tamil Nadu 601301.

Frequently Asked Questions

What does Respiratory and Aerosol Device Testing cover? +

It covers six checkpoints — connector integrity, airflow resistance, leakage, pressure drop, nebulizer output, and aerosol particle size distribution — referenced to EN 13544-1, ISO 23328-1, ISO 80601-2-12, and USP <601>.

Why does airflow resistance testing for medical device circuits use a breathing simulator head? +

A calibrated simulator reproduces realistic inhalation and exhalation flow rates, so the resistance measured reflects what a compromised patient would actually experience rather than a static bench reading at a single arbitrary flow.

How is this checkpoint actually measured? +

The device is run for a fixed duration and the aerosolized medication mass collected downstream is weighed, giving a delivered-dose-per-minute figure that a clinician can compare directly against the prescribed dose.

Why does this checkpoint matter so much? +

Particles outside the respirable size range simply deposit in the mouth and throat rather than reaching the lower airway, so a nebulizer can pass every other check and still fail to deliver an effective dose if its droplet size is wrong.

How do I get a quote for this testing? +

Call us at 90876 86986, message us on WhatsApp, or fill in the enquiry form below with your device and testing needs.

Book Your Respiratory Device Testing Today

Talk to our Oragadam team about airflow resistance, nebulizer output, or a full respiratory device study for your medical device testing programme.

Call 90876 86986 WhatsApp Us

Request Respiratory Device Testing

Fill in your details below and our team will get back to you shortly.

May i Help You!
Scroll to Top